Miris Calibration Control Kit consists of two different solutions: Miris Calibration Control 1 and Miris Calibration Control 2.These are standardized solutions with known concentrations of fat, protein and carbohydrates that correspond to those found in average human milk and the high end of the Miris HMA® measuring range.
Application: Only for validation of the preset internal calibration of the Miris HMA®.
Storage: Store the concentrate dark, at room temperature (20-30°C, 68-86°F), sealed in container in which it is supplied. Store the diluted solution in glass or plastic containers, out of direct sunlight. Keep away from foodstuff and animal feed.
Instructions: Dilute 10 ml concentrate (1 tube) with 90 ml distilled or deionized water and mix. Warm to 40°C (104°F) before use. The estimated time for preparation is 15 minutes for mixing the ingredients and to warm the solution to 40°C (104°F). The solution is to be used when preparing the instrument for analysis and also after cleaning if analysis is continued.
Application: Only for cleaning of the Miris HMA® cuvette and surface of the instrument. For in vitro diagnostic use.
Storage: Store the concentrate dark, at room temperature (20-30°C), sealed in the contain- er in which it is supplied. Store the diluted solution in glass or plastic containers, out of direct sunlight. Keep away from foodstuff and animal feed.
Instructions: Dilute 50 ml concentrate (1 tube) with 950 ml distilled or deionized water and mix. Warm to 40° C before use and follow the instructions in Chapter 3.
A 5ml tube that fits in the tube holder supplied with Miris Ultrasonic Processor™. The tubes can be used throughout the Milk Analysis process ea. storage, warming and homogenization.
A freestanding 10ml tube with a screw lid tube that fits in the tube holder supplied with Miris Ultrasonic Processor™. The tubes can be used throughout the Milk Analysis process ea. storage, warming and homogenization.
Syringes are used to inject milk samples and solutions in to the Miris HMA®.
Syringes are used to inject milk samples and solutions in to the Miris HMA®.
There was an error trying to send your message.
Please try again later.
Or send as an e-mail to info@mirissolutions.com.
There was an error trying to send your message.
Please try again later.
Or send as an e-mail to order@mirissolutions.com.
There was an error trying to send your message.
Please try again later.
Or send as an e-mail to support@mirissolutions.com.